Comparing Unrelated Readings
Do not calculate polarization change from different structures, different test points, or readings taken under incompatible conditions.
Polarization change calculations compare related structure-to-electrolyte potentials to determine how much a structure has polarized or depolarized. These calculations are commonly used when evaluating the 100 mV polarization criterion.
The calculation is based on the magnitude of the potential shift. The direction of the shift matters technically, but the comparison is usually reported as a positive millivolt change.
Polarization Change = absolute value of (Potential 1 - Potential 2)
| Term | Meaning | Common Unit |
|---|---|---|
| Potential 1 | Native, polarized, instant-off, or depolarized potential, depending on the test method | mVCSE |
| Potential 2 | Comparable potential measured at the same structure location and reference scale | mVCSE |
| Polarization Change | Magnitude of the potential shift | mV |
The two readings must be comparable. They should represent the same structure, same test location, same reference electrode scale, and valid test conditions.
A structure has a native potential of −650 mVCSE. After CP is applied, the polarized potential is −780 mVCSE. Determine the polarization shift.
Polarization Change = absolute value of [-780 - (-650)]
Polarization Change = absolute value of [-130]
Polarization Change = 130 mV
The structure polarized by 130 mV at that location. If the applicable requirement allows the 100 mV polarization criterion and the data are valid, this shift would satisfy a 100 mV polarization requirement at that test location.
A structure has an instant-off potential of −820 mVCSE. After depolarization, the potential is −700 mVCSE. Determine the depolarization.
Depolarization = absolute value of [-820 - (-700)]
Depolarization = absolute value of [-120]
Depolarization = 120 mV
The measured depolarization is 120 mV. This supports a 100 mV depolarization result if the current sources were properly interrupted or removed and the timing and measurement conditions were valid.
A structure has an instant-off potential of −760 mVCSE. After depolarization, the potential is −690 mVCSE. Determine whether the measured depolarization satisfies a 100 mV polarization criterion.
Depolarization = absolute value of [-760 - (-690)]
Depolarization = absolute value of [-70]
Depolarization = 70 mV
The measured depolarization is 70 mV. Based on these values alone, the 100 mV polarization criterion is not satisfied at that test location.
Do not calculate polarization change from different structures, different test points, or readings taken under incompatible conditions.
Depolarization data may be invalid if another CP current source remains active and continues to influence the structure.
Do not compare mVCSE, zinc, and silver-silver chloride readings as though they are the same reference scale.